Reports 1-1 of 1 Clear search Modify search
CRY (Cryostat EX)
takahiro.miyamoto - 16:53 Friday 06 July 2018 (5476) Print this report
ETMX warm up work

[ Takahashi(ICRR), Miyamoto]

 

We injected the dry air again. We measured the velocity of this injection. It is ~3000Pa/h.

From this results, we think that it takes ~26 hours for returning to air pressure.

Now pressure is 2.9*10^4Pa, and payload temperature is ~240K.

Comments to this report:
takahiro.miyamoto - 17:35 Friday 06 July 2018 (5477) Print this report

I made the time - pressure graph.

Images attached to this comment