Reports 1-1 of 1 Clear search Modify search
IOO (IMC)
stefan.ballmer - 17:10 Wednesday 20 June 2018 (5329) Print this report
Removed 0.8 Hz pole in PMC MIXER and FAST path in TTFSS Interface board (D040423, S1503961)
Removed Capacitors C52 and C53 in PMC TTFSS Interface board (D040423, S1503961). This removes the 0.8Hz pole in the monitor read-back for the MIXER and FAST channel.

Additionally, we are currently almost saturating the PZT_SLOW output path. We could fix that by pulling Capacitor C41 on (D040423, S1503961), removing the 16Hz pole. But that would increase the DAC noise reaching the drive signal. We didn't pull C41 yet, because we are hoping to instead lower the frequency noise around 100Hz.

Attached are also a number of pictures of the PMC TTFSS Interface chassis. It consists of the bollowing boards:
- TTFSS Interface (D040423, S1503961, SN002)
- Low Noise Power Module Board (D0901846 rev D, S1504402, SN004)
- TTFSSI Power (JGW-D1301832-vs, S1604815, SN012

(Also see TTFSS Interface Top Assembly JGW-D1301752).
Images attached to this report
Search Help
×

Warning

×