AOS (General)
satoru.takano - 13:56 Wednesday 23 May 2018 (5046) Print this report
calibration of LVDT

[Takano, Miyazaki, Akutsu]

In Mitaka (22 May, 2018)

We proceeded calibration of LVDT coils for TMS stages just as report 5022.

1. At first we accidentially changed the gain of preamp of coil No.1, so we had to calibrate it again. 

2. Next we calibrated 2 coils, No.2 and No.3. We measured outputs at some points on the axis and a bit far from axis.

Results:

On the axis

  • No.1 : 6.48(3)e3(count/mm) * x(mm) + 0.88(6)e3(count)
  • No.2 : 6.85(4)e3(count/mm) * x(mm) - 1.11(9)e3(count)
  • No.3 : 6.65(4)e3(count/mm) * x(mm) - 0.04(9)e2(count)

Off the axis (at x = 0 mm)

  • No.2 : -32(3) (count/mm) * r(mm) - 943(5) (count), coupling ratio = 0.46%
  • No.3 : -24(2) (count/mm) * r(mm) - 210(3) (count), coupling ratio = 0.37%

You can get these data here: https://www.dropbox.com/sh/jfapmnn1l4oxn49/AABxH0C3KQFrOlIAHD1Z09ypa?dl=0

Images attached to this report
Comments to this report:
satoru.takano - 21:54 Saturday 02 June 2018 (5158) Print this report

I forgot to comment this:

The actual tangent of coil No.2 is inverse sign because we accidentially connected pins of the coil the other way.