VIS (General)
ayaka.shoda - 13:06 Tuesday 24 April 2018 (4807) Print this report
Mismtach between BF GAS and IMV OSEM

BF GAS and IMV OSEM motion does not match, even though two sensors monitors mechanically same at lower frequencies than the resonant frequency of IRM V which is more than 10 Hz.

BF GAS and IMV OSEM differs by

BS: BF GAS response is bigger by 2.3 times than IMV OSEMs.
PR2: BF GAS response is bigger by 3.3 times than IMV OSEMs.
PR3: To be investigated
PRM: almost the same (0.3 dB difference)
SR2: almost the same (0.4 dB difference)

Comments to this report:
ayaka.shoda - 14:13 Thursday 17 May 2018 (5030) Print this report

PR3: BF GAS response is bigger by 2.3 times than IMV OSEMs.

yuya.arai - 11:46 Thursday 24 May 2018 (5061) Print this report

PRM: I measured Transfer Function of PRM today, and I found that BF GAS of PRM is also bigger than by about 2 times (not exactly 2.3) than IM V.(please see the figure)

I don't know from when its value is different.

Images attached to this comment