VIS (SR3)
enzo.tapia - 14:11 Friday 06 April 2018 (4650) Print this report
SR3 BF GAS and SF GAS calibration

On Thursday 05.04 with Fujii-san, Tanioka-san and Kozu-kun:

We completed the calibration of the BF_GAS lvdt and the SF_GAS lvdt for the SR3.
We completed the calibration of the BF_GAS lvdt and the SF_GAS lvdt for the SR3.
 
For SR3_BF (BF_No7):
  • LVDT reading with the keystone at the highest position: 32767[cnt]
  • LVDT reading with the keystone at the lowest position: -29965[cnt]
  • Calibration factor: 1/(18190.5) [cnt/mm]=0.05497 [um/cnt]
  • Offset: 5170[cnt]

Details on klog 4644

http://klog.icrr.u-tokyo.ac.jp/osl/?r=4644

 

For SR3_SF (SF-B2-6):

  • LVDT reading with the keystone at the highest position: 32767[cnt]
  • LVDT reading with the keystone at the lowest position: -28665[cnt]
  • Calibration factor: 1/(11305.45) [cnt/mm]=0.08845 [um/cnt]
  • Offset: 1690 [cnt]

I have now set the calibration factors and the offset into the digital system and I took a Snapshoot.

Comments to this report:
enzo.tapia - 13:46 Monday 09 April 2018 (4656) Print this report
I realized I didn't add the graph for the SF LVDT calibration.
I realized I didn't add the graph for the SF LVDT calibration.
I realized I didn't add the graph for the SF LVDT calibration.
Images attached to this comment