[Y.Yamamoto, T.Miyamoto, K.Hasegawa]
We tried debugging of cryocon DAQ system.
What was the problem?
When the ethernet connection did not succeed, the script to get data from cryocon was aborted and data were not corrected.
What we did.
We replaced python script file to new one which yamamoto-san kindly made.
This script can deal with above connection error, it return ”-1" when it find connection error.
Furthermore, I prepared remote rebooting script for cryocon.
You can find this script in below directory :
/opt/rtcds/userapps/release/cry/k1/scripts/cryocon_rb.py
Now you can find cryocon reboot button on the cryocon medm screens(Fig.1).
When you push that button, new screen come up and you can put the ch name which you want to reboot(Fig.2).
Problem
We found the mistake in new script file.
Because of this mistake, the corrected data for below channels between about 2018/4/2 11:00 ~ 16:30 had wrong.
If you want to fix this problem, plase replace the channel name as below for this period (information only for important CH) :
K1:AOS-TEMPERATURE_IY_WAB_TEST1 -> K1:CRY-TEMPERATURE_IY_PL_PF
K1:AOS-TEMPERATURE_IY_WAB_TEST2 -> K1:AOS-TEMPERATURE_IY_WAB_TEST1
K1:AOS-TEMPERATURE_IY_WAB_TEST3 -> K1:AOS-TEMPERATURE_IY_WAB_TEST2
K1:AOS-TEMPERATURE_IY_WAB_TEST4 -> K1:AOS-TEMPERATURE_IY_WAB_TEST3
K1:AOS-TEMPERATURE_IY_WAB_TEST5 -> K1:AOS-TEMPERATURE_IY_WAB_TEST4
K1:AOS-TEMPERATURE_IY_WAB_TEST6 -> K1:AOS-TEMPERATURE_IY_WAB_TEST5
K1:AOS-TEMPERATURE_IY_WAB_TEST7 -> K1:AOS-TEMPERATURE_IY_WAB_TEST6
K1:AOS-TEMPERATURE_IY_WAB_TEST8 -> K1:AOS-TEMPERATURE_IY_WAB_TEST7
K1:AOS-TEMPERATURE_IY_WAB_TEST8 -> K1:AOS-TEMPERATURE_IY_WAB_TEST7