Bottom filter magnetic actuation calibration.

With Mark and Kozu-kun.

We measured the keystone displacement (in units of µm) as a function of the coil actuation (in units of counts). We investigated the range from -20,000 to 20,000 counts, which moved the keystone from approximatelly -1 mm to 1 mm.

The result is 20.37  counts/µm.