The real time software issue was solved, thanks to Yamamoto-kun and the CDS test bench!
We found that there was VERY strange technical rules when two arc-tangents are used in one model. The first the model needs OSC (stragne!), and the second point is that any filter banks cannot be used in the same layer of the arc-tangent part (cdsFunctionCall). Very strange.
Anyway, now two arc-tangents run in one model. So only one model; k1vists is running on k1imc0 machine now.