[Ishizaki, Okutomi]
All signals obtained
- We added the adapter cable for the BF LVDTs
- Now all the sensor signals are obtained successfully at the digital system
Recovery from black out
- Turning on the local oscillator for LVDT modulation
- Resetting the accelerometer control (because they had got oscillated)
Setup for remote operation of stepper motors
- 30m LAN cables for the stepper motor driver are laid
-
We set the stepper motor drivers available via the network
- Last week we reset the settings of the stepper motor drivers for an immediate operation of the IP's motors
- Now the two drivers are changed from real COM mode to TCP server mode
-
IP address
- K1:STEPPER_EXV_GAS: 10.68.150.45
- K1:STEPPER_EXV_IP: 10.68.150.46
- More details >>> KAGRA wiki
- Operation check hasn't done yet with the TCP server mode