IMMT1 mirror set in suspension

[Ohishi, Doi]

19th July 2017

IMMT1 mirror was moved into the vacuum chamber and was put in its suspension.
We have not hung the mirror yet.

Two problems were found during test assembly of new and modified mechanical parts;

*One EQ stop screw was a little too short to fix the mirror;
 It works as EQ stop so we decided to use it as it is.

*Coil bobbins and coil support plates interfere about 1mm.
 We asked Takamura-san (University of Toyama, Faculty of Engineering)
 to shave the plate a few mm.
 