[Hirata, Ikeda-san, Washimi-san]
We apply first contact on AR and HR side of PR3 mirror. Unfortunately, after peeling off FC, some edge and tag foot print was left.
FC procedure is as below.
1.Lock Suspension(IM→RM→TM)
2.Cleaning(floor, inside chamber by vectra-α with ethanol and vacuum cleaner)
3.Surface inspection of HR and AR side by photo
4.Set FC jig
5.FC apply(AR→HR) 40min~1h each side
6.wait 2h30min, and pill off
7.Surface inspection by photo
Many dust which we found before FC was removed. But we found some FC residue. I talked with Miyoki-san, and tomorrow we will partialy apply FC to PR3 again.
I uploaded pictures to KAGRA dropbox here.