kentaro.komori - 15:14 Thursday 19 January 2017 (2359) Print this report
Six OSEMs calibration : 033-038

[Aritomi,Shimoda,Komori]

 

At Mitaka: We calibrated S/N 033~038 OSEMs. 

The results are as follows.

 

S/N   Maximum [V]   Gain [V/mm]

033      4.86       4.43±0.02

034      4.39       4.11±0.04

035      5.19       4.53±0.02

036      4.24       4.09±0.02

037      4.89       4.45±0.05

038      4.78       4.09±0.03

 

A graph of the output signal on graduations of micrometer is attached.

There are no significant problems.

 

The variation of these OSEMs is roughly around 10%.