Reports 1-1 of 1 Clear search Modify search
VIS (IX)
takaaki.yokozawa - 18:23 Monday 12 July 2021 (17482) Print this report
Re-calibration of the IX TM length oplev
[Ushiba, Kasamatsu, Chiba, Yokozawa]

Because the measured point of the fitting range was too small in previous calibration,
we tried to do calibration for IX TM length oplev again.

The result showed
10.06[cnt/mm] as described in Fig.1.
Images attached to this report
Comments to this report:
takafumi.ushiba - 9:18 Tuesday 13 July 2021 (17488) Print this report

According to this result and optical layout measurement reesults in klog17461, new calibration factor is 439.5 um/cnt.
This value is 1-2% larger than before but not so drastic change.