With Hirata-san.
Pictures at SR3 remedying work after O3.
Hirata-san re-assembled the thermistor compond with a few changes:
- He added kapton sheet 12.5 µm thick between the thermistor and the aluminium holder.
- He covered the legs of the thermistor with PEEK tube.
- Then he measured the resistance at the in-vacuum male connector between pins 1 and 6: 111.0 Ω which is similar to previously measured values.
- He re-installed the assembly in the security structure of the suspension.