Reports 1-1 of 1 Clear search Modify search
VIS (PR3)
fabian.arellano - 17:47 Wednesday 21 April 2021 (16533) Print this report
Comment to Thermistor installation (16520)

With Hirata-san.

Pictures at SR3 remedying work after O3.

Hirata-san re-assembled the thermistor compond with a few changes:

  • He added kapton sheet 12.5 µm thick between the thermistor and the aluminium holder.
  • He covered the legs of the thermistor with PEEK tube.
  • Then he measured the resistance at the in-vacuum male connector between pins 1 and 6: 111.0 Ω which is similar to previously measured values.
  • He re-installed the assembly in the security structure of the suspension.