IOO (General)takahiro.yamamoto - 10:12 Thursday 11 March 2021 (16227)
Print this reportLaser shutter driverPresent shutter implementation seems to be not safe.
We need to modify the implementation of beam shutter ASAP and we might have to stop using it until modification.
Comments to this report:
takahiro.yamamoto - 23:25 Thursday 11 March 2021 (16241)
Print this reportSpare shutter switch was able to be driven by BIO on the test stand. Fig1 and Fig2 show the turn ON (== 1) and turn OFF (== 0), respectively.
----- Setup: BIO: using the 24th bit on the test model. BIO-Dsub converter: with +18V, JP24 is connected. Switch box: with +/- 24V
In the case that JP24 is opened, shutter switch cannot be driven.