Dsub-BNC converter for laser shutter doesn't work well
From Friday, we cannot open the laser shutter by DAC output.
I found there is not output voltage on BNC port because jumper for GND of circuit board and BNC connector.

10V had been able to be seen on BNC output by inserting jumper.
Because beam shutter surely worked during O3GK, GND level of shutter switch circuit may changed in recent days.