takahiro.yamamoto - 19:46 Friday 05 March 2021 (16172) Print this report
PMC lock was recovered
[Akutsu, Oshino, Yamamoto, with Kenta (remote)]

PMC lock was recovered with guardian.
But we suspended to output laser to IMC today because of the trouble on DGS workstation.


-----
At first, we tried to lock PMC with IO guardian. But it didn't works well. (stopped READY state.)
After then, we can lock PMC with unmanned-mode of PMC guardian.

For reducing input power of IMC, we try to use HWP at PSL
but it cannot be used because of the trouble on the DGS workstation.
- Network for k1ctr12 must be recovered.
- k1ctr12 cannot be booted up with HDMI-VGA connector. (I unplugged it.)
- One USB extender is maybe broken. (We cannot use mouse and keyboard at the same time on k1ctr12.)
Comments to this report:
takahiro.yamamoto - 1:05 Saturday 06 March 2021 (16175) Print this report
I requested DOWN for PMC guardian because temperature servo is not stable.
Until cooling down PMC, I'll keep down state.