Reports 1-1 of 1 Clear search Modify search
CAL (General)
dan.chen - 18:13 Sunday 31 January 2021 (15870) Print this report
YPcal noise investigation
Dan Chen, Darkhan Tuyenbayev
2021/1/29

We measured YPcal noise.

Guesses or results obtained:
1. We could not observe big oscillation.
2. Observed small oscillations(57MHz, 83kHz) were not reduced by our special connectors.
3. Electrical noise of YPcal is very small.
4. The original laser noise of YPcal seems large.

Detailed report: https://dac.icrr.u-tokyo.ac.jp/KAGRA/DAWG/CAL/PCAL/exp_KAGRA/210129
Images attached to this report