I began measuring the decay times of the supension.
The files are in the following directory:
/kagra/Dropbox/Subsystems/VIS/TypeBData/SR2/quality_factors/
I began measuring the decay times of the supension.
The files are in the following directory:
/kagra/Dropbox/Subsystems/VIS/TypeBData/SR2/quality_factors/
Today I continued with the measurements.
Pending review of the quality of the data which may suggest repeating a measurement, I have finished the measurements with the control system on.
I still need to do the measurements with the control system off.
Today I continued with the measurements of the decay times. I turned the control system off today.
Today I continued the day time measurements.
Today I continued the decay time measurements.
In case you do't see the plots press here: 14510.
(Code referred to below is commented and the details can be found there.)
The envelope of the decaying oscilation was calculated using the Hilbert Transform in Matlab and then a decaying exponential was fitted to the envelope. In general, this method worked becuase in most of the cases the amplitude of other resonant modes present were not large enough to create a problem. The plots for each case can be seen by running the script indicated below. In some cases there were oscllations at lower frequencies whose amplitudes precent me frok using the same method. In those cases I used two approcches. One approach was to fit a sinusoidal fuction to the data and then substract the fitted curve to the data. This worked in two cases in which the sinusoidal fit was very good (BF at 226 mHz and 609 mHz with control on), but it failed when the fit wasn't good due to the presence of more modes (same modes with control off). In these more challenging cases I changed the method to calculate the envelope. The envelope was made to follow the higher frequency peaks and troughs only and the decaying amplitude was then characterized by the difference between the high and low envelopes. See fourth picture.
Experimental data
Data files and scripts. The main script is decay_times_main.m
(Files haven't been moved to this location yet.)
Concerning the IP mode at 406 mHz (measured in IP-L) in which the decay time was longer with the control system on, I checked the value of the unity gain frequency. It's 85 mHz as shown in the figure.
If the anomaly comes from the control system I should try to prove it by putting a notch at 406 mHz in the control filter and then repeating the measurement.
Accordingly to the commented entry, yesterday I added a notch in the IP-L damping filter at 406 mHz. The notch has a quality factor of 10 and a depth of 40 dB.
I shoud add a notch at 671 mHz to check the behaviour.