MIF (General)
masayuki.nakano - 12:22 Tuesday 21 April 2020 (14258) Print this report
Finesse measurement

I modified the finesse measurement script. Now, the data and the plot is strage in /users/Commissioning/data/Finesse/

The results are as follows:

Xarm: 1456 +/- 21 (fail to save plot)
Yarm: 1312 +/- 26

average N = 10

Comments to this report:
tomotada.akutsu - 17:17 Tuesday 21 April 2020 (14261) Print this report

This is just a thought, but it may be possible to fit the power-reduction curve to the TRX or TRY DC signals at every lock loss during the recent observation run. Though I'm not sure how useful they would be.

masayuki.nakano - 0:25 Wednesday 22 April 2020 (14263) Print this report

Mmm, ring-down measurement at the lock-loss is complicated. We need to consider about:

  • How fast the injection beam was shutted out?
  • What is the main reason of the ring-down? IMC lock loss, kick of ETMs, PRM, BS or ...? It is difficult to say.
  • Arm cavity is not independent from the other arm cavity and the PRC. So, the finesse of each cavity and the PRC affect to the ring-down signal
  • And more

I don't know what kind of information we can extract from the ring-down of each lock loss. Do we have?