MIF (Noise Budget)
yuta.michimura - 21:41 Monday 16 March 2020 (13600) Print this report
Comment to Noise budget for PRFPMI (13481)

I have updated NB for PRFPMI on Mar 9, 2020 to include recent suspension related noises.


The NoiseBudgetter configuration file lives in
/kagra/Dropbox/Subsystems/MIF/NoiseBudget/PRFPMI/NPconf_20200316_1040.csv

Details:
* Type A controls noise
 - Local damping feedback signal calibated to DARM (basically sensor noise).
 - Sum of controls noises in klog #13590

* Type A DAC noise
 - DAC noise from actuators at TM stages of Type-A suspensions.
 - Modeled with 8e-14/freq**2
 - Sum of DAC noises in klog #13589

* BS controls noise
 - Local damping feedback signal calibated to DARM (basically sensor noise).
 - Sum of controls noises in klog #13587

Thank you Lucia and Fabian for presenting noise projections in the NoiseBudgetter format!yes