VIS (EX)
shinji.miyoki - 9:17 Thursday 02 January 2020 (12375) Print this report
EXC type-A shocked by temp change in EXC room.
According to the attached data, there seemed to be a large disturbances
of many in/out signals for F0, F1, F2, BF in EXC type-A suspension because of temp increase by 0.4-0.5 C due to LED light ON in EXC1F room.

After these disturbances, some oscillation seems to exist and its damping rate seems to be very slow with a rate of a day. We should ask VIS group about this situation is healthy or not.

If the LED lights are tuned off again, the same thing will happen on the way of temp reduction.
Images attached to this report