Reports 1-1 of 1 Clear search Modify search
IOO (General)
yuta.michimura - 21:44 Wednesday 16 March 2016 (1039) Print this report
REFL table optics installed

[Yuzurihara, Kambara, Michimura]

We have setup the REFL table. The table was fixed close to the IFI chamber and optics are installed (see attached photo). REFL optics consist from a periscope, a lens (f=300 mm), a 50:50 BS, and a DC PD(Thorlabs PDA100A). Half of the beam is currently damped (we can use this for a camera).

The output of the REFL DC PD is connected to the Anti-Alias Chassis in channel IN 9 for ADC 0.

The REFL beam from the X arm show some striped pattern (see attached photo). We are not sure where this comes from, but I guess it is from beam clipping somewhere. The DC PD at BS was already removed.

Next:
 1. Measure the power at REFL
 2. Modify LSC model to include REFL DC PD.

Images attached to this report
Comments to this report:
tomotada.akutsu - 22:29 Wednesday 16 March 2016 (1040) Print this report

> The REFL beam from the X arm show some striped pattern (see attached photo)

One thing the beam pattern reminds me is that the transmitting light through the IFI that was clipped by the IFI...

Search Help
×

Warning

×