Reports 1-1 of 1 Clear search Modify search
CAL (General)
yuki.inoue - 17:44 Wednesday 04 September 2019 (10328) Print this report
Prediction of CAL line
Member: I.Yuki, S.Takahiro, C,Wu-Cheng

To inject the Pcal line for the calibration, we estimated the Prediction of CAL line based the current measured sensitivity.
The candidate of line frequency is 15.8Hz. First, we test the line with PCAL only.
After that, we inject the software and actuator lines at the same time.

Assumption:
-20dB attenuator
-20W
-80% AOM efficiency
-97.8% optical efficiency
-DWF for DARM: ZPK((10;10)(1;1)1)
-ARM length: 3000
-unit conversion: nm